Literature DB >> 19071588

New automatized method with amperometric detection for the determination of azithromycin.

Miriam E Palomeque1, Patricia I Ortíz.   

Abstract

A FIA-amperometric method for azithromycin determination was developed. A working glassy carbon electrode and a Ag/AgCl/NaCl (3M) reference electrode were used. The determination is based on the electrochemical oxidation of the azithromycin at 0.9V in Britton-Robinson buffer solution (pH 8.0). Due to the adsorption of the reaction products on the electrode surface, an effective cleaner cycle was implemented. By using the optimum chemical and FIA conditions, a concentration linear range of 1.0-10.0mgL(-1) and a detection limit (LOD) of 0.76mgL(-1) are obtained. The method was validated and satisfactorily applied to the determination of azithromycin in pharmaceutical formulations.

Entities:  

Year:  2006        PMID: 19071588     DOI: 10.1016/j.talanta.2006.10.002

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Spectrophotometric estimation of azithromycin in tablets.

Authors:  B K Jayanna; G Nagendrappa; N Gowda
Journal:  Indian J Pharm Sci       Date:  2012-07       Impact factor: 0.975

  1 in total