| Literature DB >> 19071588 |
Miriam E Palomeque1, Patricia I Ortíz.
Abstract
A FIA-amperometric method for azithromycin determination was developed. A working glassy carbon electrode and a Ag/AgCl/NaCl (3M) reference electrode were used. The determination is based on the electrochemical oxidation of the azithromycin at 0.9V in Britton-Robinson buffer solution (pH 8.0). Due to the adsorption of the reaction products on the electrode surface, an effective cleaner cycle was implemented. By using the optimum chemical and FIA conditions, a concentration linear range of 1.0-10.0mgL(-1) and a detection limit (LOD) of 0.76mgL(-1) are obtained. The method was validated and satisfactorily applied to the determination of azithromycin in pharmaceutical formulations.Entities:
Year: 2006 PMID: 19071588 DOI: 10.1016/j.talanta.2006.10.002
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057