| Literature DB >> 19071423 |
Xin-Ai Li1, Dong-Mei Zhou, Jing-Juan Xu, Hong-Yuan Chen.
Abstract
In-channel indirect amperometric detection mode for microchip capillary electrophoresis with positive separation electric field is successfully applied to some heavy metal ions. The influences of separation voltage, detection potential, the concentration and pH value of running buffer on the response of the detector have been investigated. An optimized condition of 1200V separation voltage, -0.1V detection potential, 20mM (pH 4.46) running buffer of 2-(N-morpholino)ethanesulfonic acid (MES)+l-histidine (l-His) was selected. The results clearly showed that Pb(2+), Cd(2+), and Cu(2+) were efficiently separated within 80s in a 3.7cm long native separation PDMS/PDMS channel and successfully detected at a single carbon fibre electrode. The theoretical plate numbers of Pb(2+), Cd(2+), and Cu(2+) were 1.2x10(5), 2.5x10(5), and 1.9x10(5)m(-1), respectively. The detection limits for Pb(2+), Cd(2+), and Cu(2+) were 1.3, 3.3 and 7.4muM (S/N=3).Entities:
Year: 2006 PMID: 19071423 DOI: 10.1016/j.talanta.2006.06.009
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057