Literature DB >> 19071265

Determination of trace elements in natural waters by inductively coupled plasma time of flight mass spectrometry after flow injection preconcentration in a knotted reactor.

Boryana Dimitrova-Koleva1, Karima Benkhedda, Elisaveta Ivanova, Freddy Adams.   

Abstract

A flow injection on-line sorption system was developed for the separation and preconcentration of traces of Ag, Cd, Co, Ni, Pb, U and Y from natural water samples with subsequent detection by ICP TOF MS. Simultaneous preconcentration of the analytes was achieved by complexation with the chelating reagent 1-phenyl-3-methyl-4-benzoylpyrazol-5-one immobilized on the inner walls of a (200cm x 0.5mm) PTFE knotted reactor. The analytes were eluted and transported to an axial ICP TOF MS system with 1% (v/v) HNO(3) containing 0.3microg l(-1) of Rh as an internal standard using ultrasonic nebulization. The detection limits (3sigma) varied from 0.3ng l(-1) for Y to 15.2ngl(-1) for Ni and the precision (R.S.D.) was better than 4%. Using a loading time of 90s and a sample flow rate of 4.5ml min(-1), enhancement factors of 3-14 were obtained for the different analytes in comparison with their direct determination by ICP TOF MS with ultrasonic nebulization without preconcentration. The accuracy of the method was demonstrated by analysis of water based certified reference materials.

Entities:  

Year:  2006        PMID: 19071265     DOI: 10.1016/j.talanta.2006.03.014

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Trace lead measurement and online removal of matrix interference in geosamples by ion-exchange coupled with flow injection and hydride generation atomic fluorescence spectrometry.

Authors:  Chun-Hua Tan; Xu-Guang Huang
Journal:  J Autom Methods Manag Chem       Date:  2009-09-06
  1 in total