Variable beam dose rate and DMLC IMRT to moving body anatomy.

Lech Papiez1, Ramin M Abolfath.   

Abstract

Derivation of formulas relating leaf speeds and beam dose rates for delivering planned intensity profiles to static and moving targets in dynamic multileaf collimator (DMLC) intensity modulated radiation therapy (IMRT) is presented. The analysis of equations determining algorithms for DMLC IMRT delivery under a variable beam dose rate reveals a multitude of possible delivery strategies for a given intensity map and for any given target motion patterns. From among all equivalent delivery strategies for DMLC IMRT treatments specific subclasses of strategies can be selected to provide deliveries that are particularly suitable for clinical applications providing existing delivery devices are used. Special attention is devoted to the subclass of beam dose rate variable DMLC delivery strategies to moving body anatomy that generalize existing techniques of such deliveries in Varian DMLC irradiation methodology to static body anatomy. Few examples of deliveries from this subclass of DMLC IMRT irradiations are investigated to illustrate the principle and show practical benefits of proposed techniques.

Entities:  

Mesh:

Year:  2008        PMID: 19070217     DOI: 10.1118/1.2977822

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.

Authors:  Eric W Pepin; Huanmei Wu; Hiroki Shirato
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071