Clinical results of far-field R-wave reduction with a short tip-ring electrode.

Joris R De Groot1, Jutta M Schroeder-Tanka, Jaap Visser, Albert R Willems, Willem G De Voogt.   

Abstract

BACKGROUND: Far-field R-wave (FFRW) sensing of the atrial lead of AAI or DDD pacemakers causes incorrect mode switches and remains a problem in patients with atrial arrhythmias in whom low voltage sensing is essential. We studied a pacing electrode with a short tip-ring distance (1.1 mm). We compared our findings with recordings from a conventional electrode with a larger tip-ring distance (10 mm).
METHODS: Thirty-six consecutive patients with an indication for DDD pacing were implanted with the short tip-ring electrode. Another 23 patients received the conventional electrode. FFRW and P-wave amplitudes during pacing and intrinsic ventricular depolarization were measured at implantation. Measurements were repeated before hospital discharge and at follow-up between 10 and 14 days after implantation.
RESULTS: P-wave amplitude was slightly smaller in the short tip-ring group (2.71+/-1.04 vs 3.17+/-1.30 mV in the conventional group, respectively, P=NS). All P-waves exceeded 1.2 mV. FFRW during pacing was 0.07+/-0.05 in the short tip-ring group and 0.54+/-0.32 mV in the conventional group (P<0.001). FFRW during intrinsic rhythm was 0.08+/-0.04 and 0.55+/-0.31 mV, respectively (P<0.001). The ratio between P-wave and FFRW was 48.6+/-27.2 in the short tip-ring group and 7.3+/-4.4 in the conventional group (P<0.001). FFRW and P-wave amplitudes did not change at hospital discharge or during follow-up.
CONCLUSION: FFRW can be suppressed without compromising P-wave sensing by using a pacing electrode with a short tip-ring distance. Whether reduced FFRW amplitude results in clinical endpoints remains to be determined.

Entities:  

Mesh:

Year:  2008        PMID: 19067807     DOI: 10.1111/j.1540-8159.2008.01226.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave.

Authors:  Christof Kolb; Georg Nölker; Carsten Lennerz; Hansmartin Jetter; Verena Semmler; Klaus Pürner; Klaus-Jürgen Gutleben; Tilko Reents; Klaus Lang; Ulrich Lotze
Journal:  PLoS One       Date:  2012-06-22       Impact factor: 3.240

2.  Enhanced detection of atrial tachyarrhythmias with pacing devices by using more accurate atrial sensing.

Authors:  Sami Pakarinen; Mika Lehto; Jaap Ruiter; Willem G de Voogt
Journal:  J Interv Card Electrophysiol       Date:  2021-10-01       Impact factor: 1.759

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.