| Literature DB >> 19057374 |
Luca Morini1, Emilia Marchei, Manuela Pellegrini, Angelo Groppi, Cristiana Stramesi, Federica Vagnarelli, Oscar Garcia-Algar, Roberta Pacifici, Simona Pichini.
Abstract
A liquid chromatography tandem mass spectrometric (LC-MS/MS) method with postcolumn addition of acetonitrile for the determination of ethyl glucuronide (EtG) and ethyl sulfate (EtS) in meconium was developed and validated using pentadeuterated EtG and pentadeuterated EtS as internal standards. The analytes were extracted from the matrix by acetonitrile, concentrated by solid phase extraction, separated using a reversed-phase chromatographic column, and quantified within 9 minutes. Lower limits of quantification were 5 and 1 ng/g meconium for EtG and EtS, respectively. Calibration curves were linear from lower limits of quantifications to 500 ng/g, with a minimum r > 0.999. At 3 concentrations spanning the linear dynamic range of the assay, mean recoveries ranged between 78.7% and 96.8% for EtG and between 72.1% and 95.6% for EtS. Inaccuracy was better than 8.1%, with intra-assay and interassay imprecision better than 7.2% and 10.5%, respectively. Matrix effects (ion suppression/enhancement) were found to be negligible. The analytes of interest were stable at room temperature, at 4 degrees C, when exposed to 3 freeze-thaw cycles, and when stored at -20 degrees C for up to 6 months. This sensitive and specific method was used to assess the presence of these alcohol biomarkers in meconium samples from 2 different city cohorts.Entities:
Mesh:
Substances:
Year: 2008 PMID: 19057374 DOI: 10.1097/FTD.0b013e31818b2fd9
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681