Diagnostic accuracy of a reagent strip for assessing urinary albumin excretion in the general population.

Maria Stella Graziani1, Giovanni Gambaro, Lucilla Mantovani, Alessandro Sorio, Tewoldemedhn Yabarek, Cataldo Abaterusso, Antonio Lupo, Paolo Rizzotti.   

Abstract

BACKGROUND: Albuminuria is a sensitive marker of renal derangement and has been included in a number of studies investigating chronic kidney diseases (CKDs). This study is aimed to evaluate the diagnostic performances of a strip for measuring the albumin/creatinine ratio (ACR) in the general population and to compare it with those found in a diabetic population.
METHODS: Urine samples were obtained from 201 consecutive subjects enrolled in an epidemiological study and from 259 type 2 diabetic patients. Urine was tested for albumin and creatinine using the strip (Clinitek Microalbumin) and laboratory methods. A hundred samples were stored under various conditions to assess analyte stability.
RESULTS: In the general population, the strip test reached a 90% sensitivity and 91% specificity, considering the laboratory method as the 'gold standard', sparing >80% of subjects the laboratory tests at the expense of a 1% false negative rate and an 8% false positive rate. Regarding sensitivity and specificity, the ACR test performs very similarly in the general population and in the diabetics. The stability study showed that storage at -20 degrees C induced a significant decrease in the albumin concentration with both methods, such that 5% of the samples were re-classified in the lower ACR class. Storage at -80 degrees C for up to 12 months did not affect the measurement with both methods.
CONCLUSION: Clinitek Microalbumin strips can be used for screening purposes in the general population since they correctly classify a significant percentage of subjects, particularly those with a normal albuminuria. Storage at -80 degrees C does not affect strip results. Screening with the strip and confirming positive results with a wet chemistry method are an efficient strategy for detecting albuminuria in the general population.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19037085     DOI: 10.1093/ndt/gfn639

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  12 in total

1.  Small effective population size and genetic homogeneity in the Val Borbera isolate.

Authors:  Vincenza Colonna; Giorgio Pistis; Lorenzo Bomba; Stefano Mona; Giuseppe Matullo; Rosa Boano; Cinzia Sala; Fiammetta Viganò; Antonio Torroni; Alessandro Achilli; Baharak Hooshiar Kashani; Giovanni Malerba; Giovanni Gambaro; Nicole Soranzo; Daniela Toniolo
Journal:  Eur J Hum Genet       Date:  2012-06-20       Impact factor: 4.246

2.  Chapter 1: Definition and classification of CKD.

Authors: 
Journal:  Kidney Int Suppl (2011)       Date:  2013-01

3.  Evaluation of the URiSCAN super cassette ACR semiquantitative urine dipstick for microalbuminuria screening.

Authors:  Min-Chul Cho; Misuk Ji; So Young Kim; Wonho Choe; Woochang Lee; Sail Chun; Won-Ki Min
Journal:  J Clin Lab Anal       Date:  2014-02-27       Impact factor: 2.352

Review 4.  Paper-based assays for urine analysis.

Authors:  Eric Lepowsky; Fariba Ghaderinezhad; Stephanie Knowlton; Savas Tasoglu
Journal:  Biomicrofluidics       Date:  2017-10-17       Impact factor: 2.800

5.  Prevalence of CKD in northeastern Italy: results of the INCIPE study and comparison with NHANES.

Authors:  Giovanni Gambaro; Tewoldemedhn Yabarek; Maria Stella Graziani; Alessandro Gemelli; Cataldo Abaterusso; Anna Chiara Frigo; Nicola Marchionna; Lorenzo Citron; Luciana Bonfante; Francesco Grigoletto; Salvatore Tata; Pietro Manuel Ferraro; Angelo Legnaro; Gina Meneghel; Piero Conz; Paolo Rizzotti; Angela D'Angelo; Antonio Lupo
Journal:  Clin J Am Soc Nephrol       Date:  2010-09-02       Impact factor: 8.237

6.  Microalbuminuria in subjects with hypertension attending specialist blood pressure clinics.

Authors:  A A Alharf; S Cleland; J Webster; G T McInnes; S Padmanabhan
Journal:  J Hum Hypertens       Date:  2015-12-17       Impact factor: 3.012

7.  Diagnostic accuracy of semiquantitative point of care urine albumin to creatinine ratio and urine dipstick analysis in a primary care resource limited setting in South Africa.

Authors:  June Fabian; Jaya A George; Sean D Currin; Mwawi S Gondwe; Nokthula B Mayindi; Shingirai Chipungu; Bongekile L Khoza; Stephen Tollman
Journal:  BMC Nephrol       Date:  2021-03-20       Impact factor: 2.388

8.  Comparison of Urine Albumin-to-Creatinine Ratio (ACR) Between ACR Strip Test and Quantitative Test in Prediabetes and Diabetes.

Authors:  Eun Hee Nah; Seon Cho; Suyoung Kim; Han Ik Cho
Journal:  Ann Lab Med       Date:  2017-01       Impact factor: 3.464

9.  Can a semi-quantitative method replace the current quantitative method for the annual screening of microalbuminuria in patients with diabetes? Diagnostic accuracy and cost-saving analysis considering the potential health burden.

Authors:  Yaerim Kim; Seokwoo Park; Myung-Hee Kim; Sang Hoon Song; Won Mok Lee; Hye Soon Kim; Kyubok Jin; Seungyeup Han; Yong Chul Kim; Seung Seok Han; Hajeong Lee; Jung Pyo Lee; Kwon Wook Joo; Chun Soo Lim; Yon Su Kim; Dong Ki Kim
Journal:  PLoS One       Date:  2020-01-21       Impact factor: 3.240

10.  Impact of smoking on microalbuminuria and urinary albumin creatinine ratio in non-diabetic normotensive smokers.

Authors:  R K Gupta; R Gupta; V D Maheshwari; M Mawliya
Journal:  Indian J Nephrol       Date:  2014-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.