Second instrument tip breaks during phacoemulsification.

Fariba Nazemi1, Silvia Odorcic, Rosa Braga-Mele, David Wong.   

Abstract

BACKGROUND: Second instrument tip breaks during phacoemulsification are complications that are anecdotally recalled, yet little information exists on why and how often they occur, whether they are consistently tracked, and how they are managed. They may be an underreported, but potentially serious, complication of phacoemulsification.
METHODS: We surveyed 114 cataract surgeons in Ontario to determine reported rates of second instrument tip breaks, their management, and presumed etiology. We reviewed 4 Toronto cataract centres for incident reports, instrument sterilization processes, and purchase histories. Using scanning electron microscopy (SEM), we compared the characteristics of a broken Sweeney tip to new and used second instruments.
RESULTS: Of the 35 surgeons responding to the survey, 34% had experienced a second instrument tip break during their careers. Approximately 73% (16 cases) of the 22 cases reported were managed successfully during the procedure by the primary surgeon, 14% (3 cases) required imaging by computerized tomography or x-ray, and another 14% (3 cases) required pars plana vitrectomy for tip retrieval. Purchase histories revealed that 1 Sweeney hook was exchanged monthly, equivalent to 100 to 150 surgeries. SEM of new and used second instruments revealed signs of metal fatigue on both new and used second instruments.
INTERPRETATION: Although both physicians and hospitals lack a method for ensuring quality control of second instruments, approximately one third of cataract surgeons encounter second instrument tip breaks during the course of their careers. Although most cases are managed intraoperatively, consistent hospital tracking records and standardized instrument inspection by institutions and surgeons are needed to determine how these complications occur and to establish protocols for complication reporting and management.

Entities:  

Mesh:

Year:  2008        PMID: 19020637     DOI: 10.3129/i08-148

Source DB:  PubMed          Journal:  Can J Ophthalmol        ISSN: 0008-4182            Impact factor:   1.882


  3 in total

1.  Intraoperative fracture of phacoemulsification sleeve.

Authors:  Jennifer W H Shum; Keith S K Chan; David Wong; Kenneth K W Li
Journal:  BMC Ophthalmol       Date:  2010-11-30       Impact factor: 2.209

2.  "Phacoemulsification tip fracture and how to manage it."

Authors:  Sukkarieh Georges; Al Soueidy Amine; Bejjani Riad; Chelala Elias
Journal:  Am J Ophthalmol Case Rep       Date:  2021-03-05

3.  Intraoperative fracture of phacoemulsification tip.

Authors:  Dewang Angmo; Sudarshan K Khokhar; Anasua Ganguly
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Jan-Mar