Literature DB >> 19019703

Diagnosis of type III endoleak and endovascular treatment with aortouniiliac stent-graft.

Robert Juszkat1, Ryszard Staniszewski, Anna Zarzecka, Wacław Majewski.   

Abstract

The present report describes a case of type III endoleak from a tear in the fabric of a Zenith bifurcated stent-graft approximately 6 months after implantation. The reason of the fabric tear was unknown. The complication was successfully treated by aortouniiliac stent-graft implantation followed by creation of a femorofemoral bypass.

Mesh:

Year:  2008        PMID: 19019703     DOI: 10.1016/j.jvir.2008.09.029

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  4 in total

1.  Late type III endoleak from fabric tears of a zenith stent graft: report of a case.

Authors:  Hiroshi Banno; Hirofumi Morimae; Tsutomu Ihara; Masayoshi Kobayashi; Kiyohito Yamamoto; Kimihiro Komori
Journal:  Surg Today       Date:  2012-09-14       Impact factor: 2.549

2.  Delayed Type IIIb endoleak secondary to graft fabric tear 7 years following implantation of a Medtronic Talent endovascular aortic device: A case report and review of the literature.

Authors:  Vinay Kansal; Sudhir Nagpal
Journal:  SAGE Open Med Case Rep       Date:  2016-09-20

3.  Type III B endoleak leading to aortic rupture after endovascular repair: analysis of errors in follow up and treatment.

Authors:  Marco Leopardi; Alessia Salerno; Pietro Scarpelli; Marco Ventura
Journal:  CVIR Endovasc       Date:  2018-11-06

4.  Occurrence and outcomes of type 3 endoleaks in endovascular aortic repair within the Vascular Quality Initiative database.

Authors:  Juliet Blakeslee-Carter; Adam Beck; Emily Spangler
Journal:  BMJ Surg Interv Health Technol       Date:  2020-12
  4 in total