Pinhole versus parallel-hole collimators for parathyroid imaging: an intraindividual comparison.

Maria B Tomas1, Paul V Pugliese, Gene G Tronco, Charito Love, Christopher J Palestro, Kenneth J Nichols.   

Abstract

UNLABELLED: This study was undertaken to determine the effects of collimators on the accuracy of preoperative sestamibi parathyroid imaging of the neck.
METHODS: Forty-nine patients with primary hyperparathyroidism underwent preoperative (99m)Tc-sestamibi parathyroid imaging. The protocol included early and late pinhole and parallel-hole imaging. One experienced nuclear physician, without knowledge of other test results or final diagnoses, interpreted studies. For both pinhole and parallel-hole images, focally increased sestamibi accumulation outside the normal tracer biodistribution that persisted or increased in intensity from early to late images was interpreted as positive for a parathyroid lesion. Final diagnoses were operatively confirmed in all patients.
RESULTS: Fifty-four parathyroid lesions were resected from the 49 patients. Forty-five patients had single-gland disease. Four patients had multigland disease: 3 had 2 lesions and 1 had 3 lesions. Median lesion weight was 840 mg. Pinhole imaging was significantly more sensitive than parallel-hole imaging (89% vs. 56%; P = 0.0003) for all 54 lesions. Specificity did not significantly differ between pinhole and parallel-hole imaging (93% vs. 96%, P = 0.29). Pinhole imaging was significantly more sensitive than parallel-hole imaging for lesions above (100% vs. 68%, P = 0.003) and below (77% vs. 42%, P = 0.03) the median weight and for single-gland disease (96% vs. 67%, P = 0.001). Pinhole imaging also was more sensitive for multigland disease, although the difference was only marginally significant (55% vs. 0%, P = 0.037).
CONCLUSION: Because sensitivity is significantly higher, sestamibi parathyroid imaging of the neck should be performed with a pinhole collimator.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 19008287     DOI: 10.2967/jnmt.108.055640

Source DB:  PubMed          Journal:  J Nucl Med Technol        ISSN: 0091-4916


  3 in total

Review 1.  Imaging for primary hyperparathyroidism--an evidence-based analysis.

Authors:  Radu Mihai; Dietmar Simon; Per Hellman
Journal:  Langenbecks Arch Surg       Date:  2009-07-10       Impact factor: 3.445

2.  Comparison of five parathyroid scintigraphic protocols.

Authors:  Virpi Tunninen; Pekka Varjo; Jukka Schildt; Aapo Ahonen; Tomi Kauppinen; Irina Lisinen; Anu Holm; Hannu Eskola; Marko Seppänen
Journal:  Int J Mol Imaging       Date:  2013-01-21

Review 3.  Advances in pinhole and multi-pinhole collimators for single photon emission computed tomography imaging.

Authors:  Jalil Pirayesh Islamian; AhmadReza Azazrm; Babak Mahmoudian; Esmail Gharapapagh
Journal:  World J Nucl Med       Date:  2015 Jan-Apr
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.