Equipment failure: causes and consequences in endoscopic gynecologic surgery.

Sébastien Courdier1, Olivier Garbin, Michel Hummel, Véronique Thoma, Elizabeth Ball, Romain Favre, Arnaud Wattiez.   

Abstract

STUDY
OBJECTIVE: To determine the incidence of equipment failure in gynecologic endoscopy and investigate causes and consequences.
DESIGN: A prospective observational single-center study between January and April 2006.
SETTING: Gynecologic surgery department of a university hospital.
INTERVENTIONS: In all, 116 endoscopic interventions were included: 62 laparoscopies, 51 operative hysteroscopies, and 3 fertiloscopies. Emergency and equipment testing procedures were excluded.
MEASUREMENTS AND MAIN RESULTS: Equipment malfunctions were divided into 4 categories with regard to imaging, transmission of fluids and light, the electric circuit, and surgical instruments. We also found cases with faulty connections between elements. Factors including human error, loss of time, and actual or potential consequences were analyzed. At least 1 equipment failure was noted in 38.8% of operative procedures, 41.9% of laparoscopies, and 37.3% of hysteroscopies. Fluid, gas, and light transmission was faulty in 36.2%, surgical instruments in 29.3%, the electric circuit in 22.4%, and imaging in 12.1%. Of malfunctions, 46.6% were a result of faulty connection between 2 elements. The most common cause for concern was bipolar forceps and cables in laparoscopy (42.3%) and the assembly of small parts in hysteroscopy (47.4%). Personnel were implicated in 43% of cases (nurses in 72%, surgeons in 12%, both in 16%). One equipment failure increased the total duration of laparoscopy by 7% and of hysteroscopy by 20%. The mean delay was 5.6+/-4.0minutes by equipment failure. Of the incidences, 19% could have led to serious complications for the patient; however, no morbidity or mortality actually occurred in this series.
CONCLUSION: Equipment malfunction is common in endoscopic surgery and concerns both laparoscopy and hysteroscopy. Consequences are potentially serious. It is mandatory to identify and rectify causes of equipment failure so as to optimize the daily use of endoscopic instruments and improve patient safety. The implementation of systematic checklists is currently under evaluation.

Entities:  

Mesh:

Year:  2008        PMID: 19004670     DOI: 10.1016/j.jmig.2008.08.019

Source DB:  PubMed          Journal:  J Minim Invasive Gynecol        ISSN: 1553-4650            Impact factor:   4.137


  6 in total

1.  Characterization of device-related interruptions in minimally invasive surgery: need for intraoperative data and effective mitigation strategies.

Authors:  James J Jung; Arash Kashfi; Sahil Sharma; Teodor Grantcharov
Journal:  Surg Endosc       Date:  2019-01-28       Impact factor: 4.584

2.  Prospective cohort study on surgeons' response to equipment failure in the laparoscopic environment.

Authors:  Maurits Graafland; Willem A Bemelman; Marlies P Schijven
Journal:  Surg Endosc       Date:  2014-04-26       Impact factor: 4.584

3.  Game-based training improves the surgeon's situational awareness in the operation room: a randomized controlled trial.

Authors:  Maurits Graafland; Willem A Bemelman; Marlies P Schijven
Journal:  Surg Endosc       Date:  2017-03-09       Impact factor: 4.584

4.  Instrumentation failure during microlaryngoscopy.

Authors:  Nadia van den Berg; Mel Corbett; Eoin Cleere; Ivan Keogh
Journal:  J Surg Case Rep       Date:  2022-04-25

5.  Adverse events in affiliated hospitals of mazandaran university of medical sciences.

Authors:  Benyamin Mohseni Saravi; Hasan Siamian; Ayyob Barzegar Nezhad; Zoleleykha Asghari; Azar Kabirzadeh
Journal:  Mater Sociomed       Date:  2014-04-11

6.  Surgical flow disturbances in dedicated minimally invasive surgery suites: an observational study to assess its supposed superiority over conventional suites.

Authors:  Mathijs D Blikkendaal; Sara R C Driessen; Sharon P Rodrigues; Johann P T Rhemrev; Maddy J G H Smeets; Jenny Dankelman; John J van den Dobbelsteen; Frank Willem Jansen
Journal:  Surg Endosc       Date:  2016-05-20       Impact factor: 4.584

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.