Literature DB >> 1899984

Preliminary evaluation of a prototype tube-valve-mask ventilator for emergency artificial ventilation.

P R Giffen1, C E Hope.   

Abstract

STUDY
OBJECTIVE: The objective was to design a prototype tube-valve-mask ventilator that would permit relatively inexperienced operators to provide adequate emergency artificial ventilation, namely, adequate ventilatory volumes and a high oxygen and low carbon dioxide delivery.
DESIGN: The tube-valve-mask ventilator is powered by the exhaled air of the operator and uses a tube to act as an oxygen reservoir (1,300 mL) that is filled between breaths. Mouth-to-mouth breathing was the standard against which the tube-valve-mask ventilator and the other accepted methods of mouth-to-mask and bag-valve-mask were assessed.
SETTING: Comparison studies were conducted during simulated two-person CPR using a training mannikin equipped to measure ventilation volume and delivered oxygen and carbon dioxide concentrations. TYPE OF PARTICIPANTS: Seventeen volunteer first-year nursing students were used as operators.
INTERVENTIONS: The order in which the pairs of operators performed each of the techniques was randomized.
MEASUREMENTS AND MAIN RESULTS: The ventilation volume and the percentage of oxygen and carbon dioxide delivered by each technique were as follows (mean +/- SD): Mouth-to mouth (760 +/- 290 mL, 17 +/- 1% O2, 3.4 +/- 0.4% CO2), mouth-to-mask (910 +/- 350 mL, 41 +/- 8% O2, 2.5 +/- 0.4% CO2), bag-valve-(soft) mask (550 +/- 230 mL, 94 +/- 3% O2, 0.03 +/- 0.02% CO2), bag-valve-(rigid) mask (560 +/- 300 mL, 96 +/- 3% O2, 0.03 +/- 0.02% CO2), and tube-valve-mask (860 +/- 290 mL, 91 +/- 7% O2, 0.2 +/- 0.2% CO2).
CONCLUSION: In the hands of relatively inexperienced operators, mouth-to-mouth, mouth-to-mask, and tube-valve-mask techniques provide adequate ventilation volumes to a mannikin. This was not the case with the bag-valve-mask systems (800 mL; P = .05 by t test). Of the systems that provide adequate ventilation volume, the tube-valve-mask appears, superior in that higher oxygen and lower carbon dioxide concentrations can also be obtained (P = .05 by paired t test).

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1899984     DOI: 10.1016/s0196-0644(05)80936-9

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  1 in total

1.  Terminal uridyltransferase 7 regulates TLR4-triggered inflammation by controlling Regnase-1 mRNA uridylation and degradation.

Authors:  Chia-Ching Lin; Yi-Ru Shen; Chi-Chih Chang; Xiang-Yi Guo; Yun-Yun Young; Ting-Yu Lai; I-Shing Yu; Chih-Yuan Lee; Tsung-Hsien Chuang; Hsin-Yue Tsai; Li-Chung Hsu
Journal:  Nat Commun       Date:  2021-06-29       Impact factor: 14.919

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.