Literature DB >> 18996055

Newly detected atrial fibrillation in patients with an implantable cardioverter-defibrillator is a strong risk marker of increased mortality.

T Jared Bunch1, John D Day, Brian Olshansky, Kira Q Stolen, Christopher M Mullin.   

Abstract

BACKGROUND: Atrial fibrillation (AF) has been associated with higher rates of all-cause mortality in patients with heart failure (HF). The risk of newly detected AF in patients receiving implantable cardioverter-defibrillator (ICD) therapy is unknown.
METHODS: Newly detected AF was evaluated in all patients enrolled in the Inhibition of Unnecessary RV Pacing with AV Search Hysteresis in ICDs (INTRINSIC RV) study. The relationships between AF and endpoints (total mortality, ICD shocks, and HF hospitalizations) were analyzed retrospectively with proportional-hazards models.
RESULTS: At 108 centers, 1530 patients meeting VITALITY AVT ICD indications were followed for 12 months. Of these, 1356 (89%) had no history of AF at the time of implant. Patients with a history of AF had a higher prevalence of HF (52% vs. 36%; P <.01) and had higher rates of HF hospitalization (hazard ratio [HR] 2.14 [1.29-3.54], P <.01), death (HR 2.22 [1.26-3.92], P <.01), and any ICD shock (HR 1.75 [1.19-2.58], P <.01) compared with those with no history. AF incidence during the first 3 months of implant was available in 1317 (86%; 1170 no AF, 147 history of AF) patients. New-onset AF during the first 3 months of implant (45 of 1170, 4%) was associated with a significant increased risk of death (HR 2.86 [1.02-8.05], P = .05) but not with inappropriate ICD shock (HR 2.43 [0.87-6.75], P = .09) or HF hospitalization (HR 1.17 [0.28-4.82], P = .83).
CONCLUSION: History of AF at the time of ICD implant identifies additional risk of HF and death. Newly detected AF is associated with significantly higher rates of death. The relationship between newly detected AF and inappropriate ICD shock or HF hospitalization is uncertain and requires further study.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18996055     DOI: 10.1016/j.hrthm.2008.09.025

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  9 in total

1.  Impact of atrial fibrillation on outcome of patients with idiopathic dilated cardiomyopathy: data from the Heart Muscle Disease Registry of Trieste.

Authors:  Aneta Aleksova; Marco Merlo; Massimo Zecchin; Gastone Sabbadini; Giulia Barbati; Giancarlo Vitrella; Andrea Di Lenarda; Gianfranco Sinagra
Journal:  Clin Med Res       Date:  2010-08-03

Review 2.  Adverse Outcomes from Atrial Fibrillation;Mechanisms, Risks, and Insights Learned from Therapeutic Options.

Authors:  David L Johnson; John D Day; Srijoy Mahapatra; T Jared Bunch
Journal:  J Atr Fibrillation       Date:  2012-02-02

3.  Prognostic relevance of new onset arrhythmia and ICD shocks in primary prophylactic ICD patients.

Authors:  Thomas Kleemann; Margit Strauss; Kleopatra Kouraki; Nicolas Werner; Ralf Zahn
Journal:  Clin Res Cardiol       Date:  2019-05-13       Impact factor: 5.460

4.  Older persons with diabetes receive fewer inappropriate ICD shocks: results from the INTRINSIC RV trial.

Authors:  Chirag M Sandesara; Renee M Sullivan; Andrea M Russo; WeiWei Li; Arthur Kendig; John D Day; Christopher Mullin; Kira Stolen; Brian Olshansky
Journal:  J Cardiovasc Transl Res       Date:  2010-11-18       Impact factor: 4.132

Review 5.  Rhythm control strategies and the role of antiarrhythmic drugs in the management of atrial fibrillation: focus on clinical outcomes.

Authors:  T Jared Bunch; Bernard J Gersh
Journal:  J Gen Intern Med       Date:  2010-11-25       Impact factor: 5.128

Review 6.  Pathophysiology of concomitant atrial fibrillation and heart failure: implications for management.

Authors:  T Ben Morrison; T Jared Bunch; Bernard J Gersh
Journal:  Nat Clin Pract Cardiovasc Med       Date:  2008-12-02

Review 7.  Home Screening for Detecting Subclinical Atrial Fibrillation.

Authors:  Renato Pietro Ricci; Taya V Glotzer
Journal:  J Atr Fibrillation       Date:  2015-12-31

8.  Atrial Fibrillation Is Associated With Higher Overall Mortality in Patients With Implantable Cardioverter-Defibrillator: A Systematic Review and Meta-Analysis.

Authors:  Usman Mustafa; Parinita Dherange; Rohit Reddy; Joseph DeVillier; Jessica Chong; Alarozia Ihsan; Ryan Jones; Narendra Duddyala; Pratap Reddy; Paari Dominic
Journal:  J Am Heart Assoc       Date:  2018-11-20       Impact factor: 5.501

9.  Cost Saving Potential of an Early Detection of Atrial Fibrillation in Patients after ICD Implantation.

Authors:  Thomas Reinhold; Roberto Belke; Tino Hauser; Christian Grebmer; Carsten Lennerz; Verena Semmler; Christof Kolb
Journal:  Biomed Res Int       Date:  2018-08-14       Impact factor: 3.411

  9 in total