Literature DB >> 18987532

Radiation emission dose from patients administered 90Y-labelled radiopharmaceuticals: comparison of experimental measurements versus Monte Carlo simulation.

Sergio Lo Meo1, Gianfranco Cicoria, Giancarlo Montini, Carlo Bergamini, Francesco Campanella, Davide Pancaldi, Antonio Sabatino Panebianco, Domenico Rubello, Mario Marengo.   

Abstract

AIM: To estimate the radiation dose delivered from patients injected with yttrium-90 (Y)-labelled tiuxetan (Zevalin) to parents and the general population, comparing different techniques.
METHODS: The radiation dose delivered from a group of eight patients injected with Y-Zevalin to treat recurrent lymphoma was measured. The data obtained with the Monte Carlo simulation test were compared with the experimental measurements obtained with an ionization chamber detector and with a crystal NaI(Tl) detector.
RESULTS: A good correlation was found between the Monte Carlo simulation test and the ionization chamber detector results: the air kerma dose rate was 4.2+/-0.1 and 4.4+/-0.8 microGy/h, respectively (r=0.9, P<0.01). Moreover, more than 99.7% of the air kerma dose rate measured with the ionization chamber detector was because of the contribution of electrons, whereas the contribution of photons was less than 0.3%. In contrast, the air kerma dose rate measured with the crystal NaI(Tl) detector was significantly lower (0.76+0.12 microGy/h) in comparison with the Monte Carlo simulation test. This underestimation was related to the limited crystal NaI(Tl) detector response to low energy rates at variance with the ionization chamber detector. The effective radiation dose released by patients treated with Y-labelled tiuxetan to parents and the general population was approximately 0.1 mSv per treatment cycle.
CONCLUSION: Using the Monte Carlo model as a benchmark to compare the experimental measurements obtained by the two different detectors, we found that the ionizing chamber detector was more accurate than the crystal Na(Tl) detector for measuring the exposure radiation dose delivered from patients administered with Y-labelled radiopharmaceuticals. Moreover, the effective radiation dose released by these patients to their parents and the general population is significantly lower than the value recommended by international reports and regulations.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18987532     DOI: 10.1097/MNM.0b013e328314b895

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  1 in total

1.  Physical dosimetry and mathematical dose calculation in nuclear medicine: A comparative study.

Authors:  Ali Shabestani Monfared; Mehrangiz Amiri
Journal:  Indian J Nucl Med       Date:  2010-01