Are sounds from mechanical heart valves equal for different valve types?

Thais A L Pedersen1, Peter Johansen, J Michael Hasenkam, Reiner Koerfer, Heinrich Koertke, Hans Nygaard.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The closing 'click' sounds from mechanical heart valve prostheses are often clearly audible for patients and their relatives. These sounds have been recognized as a disturbing factor causing sleeping disturbances or social embarrassment, which patients must endure for the rest of their lives. The study aim was to determine whether the intensity of the transmitted sounds varies according to the type of valve implanted.
METHODS: Closing sounds were measured from 15 patients with ATS valves, 29 with Medtronic-Hall valves, and 40 with St. Jude Medical (SJM) valve prostheses; all prostheses were implanted in the aortic position. The sounds were recorded by a microphone placed 5 cm above the chest of the patient when placed in a supine position in a bioacoustical laboratory. The mean sound pressure levels (SPLs) from the valves were measured in dB(A).
RESULTS: The mean SPL values were 43 +/- 5 dB(A) for ATS valves, 41 +/- 4 dB(A) for Medtronic-Hall valves and 40 +/- 4 dB(A) for SJM valves (p < 0.05, ATS versus SJM).
CONCLUSION: The study results showed a statistically significant difference in SPLs among the three investigated mechanical heart valve prostheses. The highest levels were in the ATS valves, and the lowest in the SJM valves. These objectively measured SPLs must be correlated with the sound as perceived by the patient in order to determine the annoyance that they cause.

Entities:  

Mesh:

Year:  2008        PMID: 18980093

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  2 in total

1.  The significance of the analysis on scalographic pattern for detecting malfunctioning bileaflet valve with the wavelet analysis.

Authors:  Hiroshi Sugiki; Kenji Sugiki; Tomonori Ooka; Satoru Wakasa; Yasushige Shingu; Tsuyoshi Tachibana; Yoshiro Matsui
Journal:  J Artif Organs       Date:  2015-09-07       Impact factor: 1.731

2.  Comparison of quality of life in patients who underwent mechanical mitral valve replacement: star GK vs SJM.

Authors:  Jiang-Shan Huang; Ning Xu; Kai-Peng Sun; Zhi-Nuan Hong; Liang-Wan Chen; Yur-Ren Kuo; Qiang Chen
Journal:  J Cardiothorac Surg       Date:  2020-01-07       Impact factor: 1.637

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.