| Literature DB >> 18970870 |
F Sánchez Rojas1, C Bosch Ojeda, J M Cano Pavón.
Abstract
A flow injection (FI) system was used to develop an efficient on-line sorbent extraction preconcentration system for palladium by graphite furnace atomic absorption spectrometry (GFAAS). The investigated metal was preconcentrated on a microcolumn packed with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide immobilized on silica gel (DPTH-gel). The palladium is eluted with 40mul of HCl 4M and directly introduced into the graphite furnace. The detection limit for palladium under the optimum conditions was 0.4ngml(-1). This procedure was employed to determine palladium in different samples.Entities:
Year: 2006 PMID: 18970870 DOI: 10.1016/j.talanta.2006.05.048
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057