Literature DB >> 18970775

Study on pressurizing electroosmosis pump for chromatographic separation.

Lei Wang1, You-Zhao He, Guo-Ni Fu, Yan-Yun Hu, Xiao-Kui Wang.   

Abstract

A porous core electroosmosis pump was studied and improved in accordance with the electroosmosis theory. Hexamethylene tetraamine (HMTA) was used as the additive of pump carrier solution to improve the flow rate stability and delivery efficiency. The influences of the electric field strength, porous core dimension and acetonitrile concentration of carrier solution on the pump flow rate and output pressure were investigated in detail. The improved electroosmosis pump can provide not only steady flow rate and large flow range, but also moderate output pressure. With the pump carrier solution of 0.5mmol/L HMTA and the working voltage of 4950V, the pump output pressure, flow rate and delivery efficiency achieved 1.1MPa, 1.3mL/min and 3.2mL/(minmA), respectively. The pump can be employed for mobile phase delivery in the reversed-phase chromatographic separation of monolithic silica columns.

Entities:  

Year:  2006        PMID: 18970775     DOI: 10.1016/j.talanta.2006.02.058

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Ion exchange resin bead decoupled high-pressure electroosmotic pump.

Authors:  Bingcheng Yang; Feifang Zhang; Xinmiao Liang; Purnendu K Dasgupta; Shaorong Liu
Journal:  Anal Chem       Date:  2009-06-15       Impact factor: 6.986

  1 in total