Hizuru Nakajima1, Maiko Yagi, Yuki Kudo, Tatsuro Nakagama, Takuya Shimosaka, Katsumi Uchiyama.
Abstract
A flow-based enzyme-linked immunosorbent assay (ELISA) on a polydimethylsiloxane (PDMS) microchip has been developed for the rapid determination of immunoglobulin A (IgA). The analytical principle of this integrated method is the same as the conventional sandwich-type ELISA. A primary antibody (anti-IgA) was adsorbed on the surface of a PDMS microchannel, and then an antigen (IgA) and a secondary antibody (anti-IgA HRP labeled) were reacted successively. The resulting antigen-antibody complex, fixed on the surface of the microchannel, was detected using Amplex((R)) Red and a fluorescent imaging system. The calibration curve of the IgA standard solution was linear in the range of 0-50ng/mL at the flow rate of 10muL/min. This flow rate corresponds to the reaction time of 4.8s. Compared to the conventional assay on a 96-well microtiter plate, the present assay on the microchip dramatically shortened the reaction time necessary for the enzyme-substrate reaction from 30min to 4.8s, i.e., to 1/375. The amounts of the reagent and sample were also reduced to 1/100 compared to the 96-well microtiter plate.Entities:
Year: 2006 PMID: 18970739 DOI: 10.1016/j.talanta.2005.12.059
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057