Literature DB >> 18970593

Development of an on-line preconcentration system for zinc determination in biological samples.

Rosilene L Dutra1, Heloisa F Maltez, Eduardo Carasek.   

Abstract

An on-line preconcentration system for zinc determination in 24-h urine, blood plasma and erythrocyte matrices by flame atomic absorption spectrometry (FAAS) was used. This procedure was based on adsorption of Zn(II) ions onto a minicolumn filled with silica gel, chemically modified with niobium(V) oxide (Nb(2)O(5)-SiO(2)). The determination of the optimum conditions for Zn(II) preconcentration was done using two-level full factorial and Doehlert designs. In the optimization procedure, four variables (sample pH, eluent concentration, sample flow rate and eluent flow rate) were investigated. The results obtained from the full factorial design demonstrated that the sample pH and sample flow rate variables, and their interactions, were statistically significant. A Doehlert matrix was used in order to determine the optimum conditions for the sample pH and sample flow rate. The optimized conditions for sample pH and flow rate sampling were 6.6 and 7.1 ml min(-1), respectively, to obtain the maximum Zn(II) preconcentration and determination in the biological samples studied. Parameters of analytical curve, precision, effect of other ions in the proposed system and accuracy were achieved to assess the proposed method. The accuracy was confirmed by analysis of certified reference materials (urine Seronorm Trace Elements) and recovery tests in blood plasma and erythrocyte samples. Detection limit (3sigma/S) of 0.77 microg l(-1), precision (calculated as relative standard deviation) of 1.5% for Zn(II) concentration of 10 microg l(-1) (n=7) and a sampling frequency of 27 samples/h were obtained from the proposed system.

Entities:  

Year:  2005        PMID: 18970593     DOI: 10.1016/j.talanta.2005.10.019

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Application of Doehlert experimental design for the removal of radium from aqueous solution by cross-linked phenoxycalix[4]pyrrole-polymer using Ba(II) as a model.

Authors:  Nancy AlHaddad; Radwan Sidaoui; Malek Tabbal; Ismail Abbas; Pierre-Edouard Danjou; Francine Cazier-Dennin; Rana Baydoun; Omar El Samad; Ahmad Rifai
Journal:  Environ Sci Pollut Res Int       Date:  2019-12-12       Impact factor: 4.223

2.  Determination of trace silver in water samples by online column preconcentration flame atomic absorption spectrometry using termite digestion product.

Authors:  Joyce Nunes Bianchin; Edmar Martendal; Eduardo Carasek
Journal:  J Autom Methods Manag Chem       Date:  2011-06-28
  2 in total