Sebastien Herrmann1, Boaz Leshem, Shimi Landes, Bracha Rager-Zisman, Robert S Marks.
Abstract
An ELISA-based optical fiber methodology developed for the detection of anti-West Nile virus IgG antibodies in serum was compared to standard colorimetric and chemiluminescent ELISA based on microtiter plates. Colorimetric ELISA was the least sensitive, especially at high titer dilutions. The fiber-optic immunosensor based on the same ELISA immunological rationale was the most sensitive technique.Entities:
Year: 2004 PMID: 18969954 DOI: 10.1016/j.talanta.2004.10.013
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057