Y Udnan1, J Jakmunee, S Jayasavati, G D Christian, R E Synovec, K Grudpan.
Abstract
A new flow injection procedure for an assay of Fe(III) by using salicylate obtained from antipyretic powder, which is a cheap and easily available reagent, is proposed. A red complex was continuously monitored by a laboratory-made green LED colorimeter. A linear calibration was obtained in the range of 1-20mgFel(-1) with a detection limit of 0.5mgFel(-1) and R.S.D.s of 1.4-5.4% (n=3, for 1-20mgFel(-1)). The new procedure was applied to assay iron contents in pharmaceutical preparations. The results were in good agreement with those of the USP standard method.Entities:
Year: 2004 PMID: 18969735 DOI: 10.1016/j.talanta.2004.03.067
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057