Sequential injection analysis is still dominated by single component analysis. In this proposed robust, economical simple instrumental system seven different metal ions are determined simultaneously using thin-film sequential injection extraction (SIE) with multivariate calibration and multiwavelength detection. Dithizone, in ethanol, is used as extractant and the metal dithizonates' spectra are generated by a diode array spectrophotometer between 300 and 700nm. The SI thin-film extraction using water miscible with ethanol works due to the hydrophobic interaction of ethanol with the Teflon wall to create a thin film. A sample frequency of 27 samples per hour was obtained with a sample carry-over of less than 1%. The results of the proposed sequential injection extraction system compare favourably with the results obtained by using standard atomic absorption spectrometry (AAS) methods on conventional extraction samples.Entities:
Year: 2004 PMID: 18969730 DOI: 10.1016/j.talanta.2004.06.020
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057