| Literature DB >> 18969380 |
X Z Wu1, P Liu, Q S Pu, Q Y Sun, Z X Su.
Abstract
A amino-terminated G 4.0 dendrimer-like polyamidoamine (PAMAM) immobilized silica gel (PAMAMSG) was prepared with a divergent method by repeating two processes: (1) Michael addition of methyl acrylate (MA) to surface amino groups; and (2) amidation of the resulting esters with ethylenediamine (EDA) from gamma-aiminopropyl silica gel (APSG) core. It was then used for the first time as microcolumn packing for the flame atomic absorption spectrometry (FAAS) determination of trace or ultra trace Pd(II), after flow injection (FI) online preconcentration and separation process. A limit of detection (LOD) of 3.9ngml(-1) was achieved when 0.200mugml(-1) Pd(II) was preconcentrated in 0.2moll(-1) HCl medium with a sampling flow rate of 6.0mlmin(-1) for 60s and the relative standard deviation (R.S.D.) was 1.7%. The proposed method was successfully used for the determination of Pd in two metallurgical samples.Entities:
Year: 2004 PMID: 18969380 DOI: 10.1016/j.talanta.2003.10.011
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057