Literature DB >> 18969025

Real-time immunoassay of ferritin using surface plasmon resonance biosensor.

Xiaoqiang Cui1, Fan Yang, Yufang Sha, Xiurong Yang.   

Abstract

A surface plasmon resonance (SPR) biosensor was used for the first time to determine the concentration of ferritin in both HBS-EP buffer and serum. The monoclonal antibody was immobilized on the carboxymethyl dextran-modified gold surface by an amine coupling method. The interaction of antibody with antigen was monitored in real-time. The signal was enhanced by sandwich amplification strategy to improve the sensitivity and specificity of the immunoassay, especially in serum. The linear range of the assay in serum is over 30-200 ng ml(-1) with the detection limit of 28 ng ml(-1). The sensitivity, specificity, and reproducibility of the assay are satisfactory. The analyte and enhancement antibody-binding surface could be regenerated by pH 2.0 glycine-HCl buffer and the same antibody-immobilized surface could be used for more than 50 cycles of ferritin binding and regeneration.

Entities:  

Year:  2003        PMID: 18969025     DOI: 10.1016/S0039-9140(03)00043-2

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Detection of human proteins using arrayed imaging reflectometry.

Authors:  Charles R Mace; Christopher C Striemer; Benjamin L Miller
Journal:  Biosens Bioelectron       Date:  2008-05-27       Impact factor: 10.618

2.  A versatile microparticle-based immunoaggregation assay for macromolecular biomarker detection and quantification.

Authors:  Haiyan Wu; Yu Han; Xi Yang; George G Chase; Qiong Tang; Chen-Jung Lee; Bin Cao; Jiang Zhe; Gang Cheng
Journal:  PLoS One       Date:  2015-02-06       Impact factor: 3.240

  2 in total