A novel analysis of explosives via the coupling of an airline passenger personnel portal with a high-flow (HF), high-resolution (HR) ion mobility spectrometry (IMS) was shown for the first time. The HF-HR-IMS utilized a novel ion aperture grid design with a (63)Ni ionization source while operating at ambient pressure in the positive ion mode at 200 degrees C. The HF-HR-IMS response characteristics of 2,4,6-trinitrotoluene (TNT), 4,6-dinitro-o-cresol (4,6DNOC), and cyclo-1,3,5-trimethylene-2,4,6-trinitramine (RDX) were investigated. Modifications made to the HF-HR-IMS exhaust and ionization source created an 800% increase in the total ion current (TIC), from 0.85 to 6.8 nA. This translated into a 65% ion response increase for TNT when compared with a traditional IMS. A mixture of TNT and (4,6DNOC) was used to successfully demonstrate the resolving power of the species with similar reduced mobility constants (K(o)), 1.54 and 1.59, respectively. The reactant ion (H(2)O)(n)H(+), peak was also used to measure the resolving power of the spectrometer while varying the internal diameter of three different aperture openings from 1.00 to 3.54cm. This provided a resolving power range of 50-60, double that typically achievable by commercial IMS instruments. Most important, these changes made in this new instrumental design can be implemented to all existing and future IMS's to greatly enhance the achievable IMS resolving power.

Entities:  

Year:  2002        PMID: 18968612

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Predicting optimal resolving power for ambient pressure ion mobility spectrometry.

Authors:  Abu B Kanu; Molly M Gribb; Herbert H Hill
Journal:  Anal Chem       Date:  2008-08-07       Impact factor: 6.986

2.  Detection of nitro-based and peroxide-based explosives by fast polarity-switchable ion mobility spectrometer with ion focusing in vicinity of Faraday detector.

Authors:  Qinghua Zhou; Liying Peng; Dandan Jiang; Xin Wang; Haiyan Wang; Haiyang Li
Journal:  Sci Rep       Date:  2015-05-29       Impact factor: 4.379

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.