Literature DB >> 18968542

Precipitation-dissolution system for silver preconcentration and determination by flow injection flame atomic absorption spectrometry.

Otoniel D Sant'ana1, Angela L R Wagener, Ricardo E Santelli, Ricardo J Cassella, Mercedes Gallego, Miguel Valcárcel.   

Abstract

In this paper, flame atomic absorption spectrometry was used for the determination of silver in various materials. The proposed preconcentration method is based on the continuous precipitation of silver as p-dimethylaminobenzilidene-rhodanine (PDBR) complex and dissolution of the precipitate with potassium cyanide. EDTA was added to the sample solution to mask large concentrations of Fe(III), Ni(II), Cu(II), Zn(II), Pb(II), Co(II) and Al(III). An enhancement factor of 20 was obtained for a preconcentration time of 3 min, resulting in a sampling frequency of 16 h(-1). The detection limit (3sigma) in the sample solution was 5 ngml(-1). The relative standard deviation at 30 ngml(-1) level was 4.7%. Analytical results obtained for alloy, biological reference material and ore samples analyzed were in good agreement with the certified values and comparable to those obtained with other techniques.

Entities:  

Year:  2002        PMID: 18968542     DOI: 10.1016/s0039-9140(01)00604-x

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Cloud point extraction-flame atomic absorption spectrometry for pre-concentration and determination of trace amounts of silver ions in water samples.

Authors:  Xiupei Yang; Zhihui Jia; Xiaocui Yang; Gu Li; Xiangjun Liao
Journal:  Saudi J Biol Sci       Date:  2017-01-23       Impact factor: 4.219

  1 in total