P A Farias1, A de Luca Rebello Wagener, A A Castro.
Abstract
An electrochemical stripping procedure for ultra-trace measurements of adenine is described based on the adsorption of the adenine-copper complex on a static mercury drop electrode. Cyclic voltammetry was used to characterize the interfacial and redox behavior. Optimum experimental conditions were found when using a 0.005 M NaOH solution containing 0.4 ppm of copper, an accumulation potential of -0.30 V, a scan rate of 100 mV s(-1) and a linear scan mode. There is a linear response to adenine concentration in the range of 0.1-1.0 ppb and the detection limit for 6 min accumulation time was of 4.0 ppt (3.0x10(-11) M). Proper conditions for measuring the adenine in presence of guanine, thymine and cytosine were also investigated. The method was applied for the determination of adenine in a sample of single-stranded calf thymus DNA.Entities:
Year: 2001 PMID: 18968371 DOI: 10.1016/s0039-9140(01)00422-2