Potentiometric sensor for hydrogen ion based on N,N'-dialkylbenzylethylenediamine neutral carrier in a poly(vinyl chloride) membrane with polyaniline solid contact.

W S Han1, M Y Park, K C Chung, D H Cho, T K Hong.   

Abstract

Hydrogen ion selective solid contact electrodes based on N,N'-dialkylbenzylethylenediamine (alkyl=butyl, hexyl, octy, decyl) are prepared. Solid contact electrodes and coated wire electrodes have been fabricated from polymer cocktail solutions based on N,N'-dialkylbenzylethylenediamine (alkyl=butyl, hexyl, octy, decyl). We showed that the response range and slopes are influenced by the alkyl chain length. Response ranges and Nernstian slops for solid contact electrodes were slightly better than for coated wire electrodes. Solid contact electrodes were shown linear selective to hydrogen ion in the pH ranges pH 4.5-13.0, 4.2-13.1, 3.4-13.0 and 3.0-13.2. and their Nernstian slopes were 49.7, 50.8, 51.5 and 53.7 mV pH(-1) at 20+/-0.2 degrees C, respectively. Stability is also improved especially when compared with coated wire electrodes. The 90% response time was <2 s and their electrical resistance varied in the range 2.37-2.76 MOmega. Especially solid contact electrodes with N,N'-didecylbenzylethylenediamine have shown the best selectivity and reproducibility of e.m.f..

Entities:  

Year:  2001        PMID: 18968237     DOI: 10.1016/s0039-9140(00)00665-2

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

Review 1.  Developments in the Field of Conducting and Non-conducting Polymer Based Potentiometric Membrane Sensors for Ions Over the Past Decade.

Authors:  Farnoush Faridbod; Parviz Norouzi; Rassoul Dinarvand; Mohammad Reza Ganjali
Journal:  Sensors (Basel)       Date:  2008-04-03       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.