Literature DB >> 18967909

Spectrofluorimetric method for determination of aluminium with chromotropic acid and its application to water samples.

C I Park1, K W Cha.   

Abstract

A rapid and sensitive method for the trace level determination of aluminium based on the formation of a 1:1 complex with chromotropic acid (1,8-dihydroxynaphthlene-3,6-disulfonic acid) in an methanol medium is reported. The fluorescence intensity of the system was 50 times greater than that of the system without aluminium. This method is very sensitive and selective for the direct determination of aluminium ion. The fluorescence is excited at 346 nm and measured at 370 nm. The optimum conditions are a chromotropic acid concentration of 5.0 ml (1.0x10(-4) M) and pH 4.0+/-0.5 (acetic acid-sodium acetate buffer). The fluorescence intensity is a linear function of the concentration of Al(III) in the range 2-100 ng ml(-1) and the detection limit is 1.0 ng ml(-1). The method has been applied successfully to the determination of trace amount of Al(III) in tap, river and sea-water samples.

Entities:  

Year:  2000        PMID: 18967909     DOI: 10.1016/s0039-9140(99)00351-3

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

1.  Online monitoring of aluminium in drinking water with fluorimetric detection.

Authors:  A Alonso-Mateos; M J Almendral-Parra; Y Curto-Serrano; F J Rodríguez-Martín
Journal:  J Fluoresc       Date:  2007-10-20       Impact factor: 2.217

2.  Thermally Stable Merocyanine Form of Photochromic Spiropyran with Aluminum Ion as a Reversible Photo-driven Sensor in Aqueous Solution.

Authors:  Jiaqiang Ren; He Tian
Journal:  Sensors (Basel)       Date:  2007-12-06       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.