| Literature DB >> 18967864 |
C Yarnitzky1, J Wang, B Tian.
Abstract
The design and operation of a hand-held electrochemical analyzer for field screening of trace lead are described. The stripping-based instrument meets the size, weight and power-consumption requirements of many decentralized lead testings, while maintaining the high sensitivity and reliability of centralized stripping operations. When coupled to screen-printed disposable sensors, the unit's response varies linearly with the lead level between 20 and 300 mug/l (50-s deposition). Comparison with lead results obtained with the small and established instruments yielded an excellent correlation (r=0.999). Convenient quantification of lead in drinking water is accomplished with quiescent, nondeaerated small samples. Such performance and portability hold great promise to a host of environmental, clinical, or industrial applications.Entities:
Year: 2000 PMID: 18967864 DOI: 10.1016/s0039-9140(99)00267-2
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057