Literature DB >> 18967756

Simultaneous determination of Cu, Cd and Pb in drinking-water using W-Coil AAS.

A Salido1, B T Jones.   

Abstract

An inexpensive, multi-element, W-coil atomic absorption spectrometer has been developed. Atomization occurs on W-coils extracted from commercially available slide projector bulbs. The system has minimal power requirements, 120 ACV and 15 A. A small, computer controlled CCD spectrometer is used as the detector. A multi-element Cu, Cd and Pb hollow cathode lamp is used as the source. 20 mul volumes are deposited on the coil and atomized at 6.7 A or approximately 2200 degrees C. Cu, Cd and Pb were simultaneously determined in tap water, drinking water and a quality control sample. The instrument detection limits are 0.8, 0.2 and 3.0 mug/l for Cu, Cd and Pb, respectively.

Entities:  

Year:  1999        PMID: 18967756     DOI: 10.1016/s0039-9140(99)00162-9

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Heavy metals pollution in a sewage treatment oxidation pond and the receiving stream of the Obafemi Awolowo University, Ile Ife, Nigeria.

Authors:  A O Ogunfowokan; A A Adenuga; N Torto; E K Okoh
Journal:  Environ Monit Assess       Date:  2007-10-24       Impact factor: 2.513

  1 in total