Literature DB >> 18967344

Immobilization of 8-hydroxyquinoline onto silicone tubing for the determination of trace elements in seawater using flow injection ICP-MS.

S N Willie1, H Tekgul, R E Sturgeon.   

Abstract

A rapid and simple on-line method is described for the preconcentration of Mn, Co, Ni, Cu, Zn, Cd and Pb from sea water using 8-hydroxyquinoline immobilized onto silicone tubing (Sil-8-HQ) via the Mannich reaction. Recoveries between 35 and 95% and limits of detection in the ppt range were obtained using a 2 m long Sil-8-HQ tube with a sample flow rate of 1.0 ml min(-1). A tube could be subjected to sample loading and elution cycles over 200 times. The capacity was 1.5 and 1.3 mug cm(-2) for Cu and Mn, respectively. Cu, Cd, Co, Pb, Mn, Zn and Ni were determined in coastal and open ocean seawater using flow injection inductively coupled plasma mass spectrometry (FI-ICP-MS). Good agreement with certified values for the certified reference materials NASS-4 and CASS-3 was demonstrated when quantitation was undertaken by the method of additions.

Entities:  

Year:  1998        PMID: 18967344     DOI: 10.1016/s0039-9140(98)00153-2

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Solid phase extraction of inorganic mercury using 5-phenylazo-8-hydroxyquinoline and determination by cold vapor atomic fluorescence spectroscopy in natural water samples.

Authors:  Mirna Daye; Baghdad Ouddane; Jalal Halwani; Mariam Hamzeh
Journal:  ScientificWorldJournal       Date:  2013-12-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.