Literature DB >> 18967005

Potentiometric pH sensors based on chemically modified electrodes with electropolymerized metal-tetraaminophthalocyanine.

T F Kang1, Z Y Xie, H Tang, G L Shen, R Q Yu.   

Abstract

Potentiometric pH sensors based on polymer film were prepared by electropolymerization of the monomer nickel(II)-4,4',4'',4'''-tetraaminophthalocyanine (NiTAPc) or copper(II)-4,4',4'',4'''-tetraaminophthalocyanine (CuTAPc) on glassy carbon (GC) electrodes. The polymer of metal tetraaminophthalocyanine (p-MTAPc) film coated electrodes show a slope of 55 +/- 1 mV/pH (at 20 degrees C) and nearly Nernstain potentiometric response to pH over the range of pH 1-13. The electrodes possess good potential reproducibility and high selectivity, and are useful sensing devices in pH determination and end-point indication of acid-base potentiometric titration.

Entities:  

Year:  1997        PMID: 18967005     DOI: 10.1016/s0039-9140(97)00124-0

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  2 in total

Review 1.  State Key Laboratory of Nonlinear Mechanics (LNM), Institute of Mechanics, Chinese Academy of Sciences, Beijing 100080, China.

Authors:  Olga Korostynska; Khalil Arshak; Edric Gill; Arousian Arshak
Journal:  Sensors (Basel)       Date:  2007-11-30       Impact factor: 3.576

2.  pH Sensitivity of Novel PANI/PVB/PS3 Composite Films.

Authors:  Edric Gill; Arousian Arshak; Khalil Arshak; Olga Korostynska
Journal:  Sensors (Basel)       Date:  2007-12-19       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.