Literature DB >> 18966991

Continuous flow analysis of lead (II) and mercury (II) with substituted diazacrown ionophore membrane electrodes.

X Yang1, D B Hibbert, P W Alexander.   

Abstract

A novel flow cell for use with ion-selective membrane electrodes is reported in which the carrier stream is drawn through a tube that suppresses the pump noise. PVC membrane electrodes based on 7,16-dithenoyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane (DTODC), and 7,16-di-(2-thiopheneacetyl)-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane (DTAODC) for lead (II), and 7,16-dithenyl-1,4,10,13-tetraoxa-7,16-diazacyclooctadecane (DTDC) and 7,16-di-(2-methylquinolyl)-1,4,10,13-tertraoxa-7,16-diazacyclootadecane (DQDC), for mercury (II) were prepared and evaluated. The linear ranges were pPb: 5.5-3.0 (DTODC) and 6.0-2.0 (DTAODC); pHg: 5.5-3.0 (DTDC) and 4.5-2.5 (DQDC). With flow rate of 3 ml min(-1) the repeatability of measurements was less than 5% RSD (n = 3). The system was applied to the determination of lead (II) and mercury (II) in spiked natural water samples.

Entities:  

Year:  1997        PMID: 18966991     DOI: 10.1016/s0039-9140(97)00122-7

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Flow injection analysis of mercury using 4-(dimethylamino) benzaldehyde-4-ethylthiosemicarbazone as the ionophore of a coated wire electrode.

Authors:  Tara F Tahir; Abdussalam Salhin; Sulaiman Ab Ghani
Journal:  Sensors (Basel)       Date:  2012-11-06       Impact factor: 3.576

  1 in total