Development of a nitrate ion-selective electrode based on an Urushi matrix membrane and its application to the direct measurement of nitrate-nitrogen in upland soils.

S Ito1, K Baba, Y Asano, H Takesako, H Wada.   

Abstract

A solid-state nitrate ion-selective electrode based on an Urushi matrix membrane was developed. Urushi, a natural oriental lacquer, has excellent mechanical strength and binding affinity for metal electrodes. Using the same technique for a dip-coating ion-selective electrode, an electrode was prepared by coating and hardening a sensing membrane on the metal base. The effects of the metal electrode on the electrode potential stability, the liquid-membrane components and the oven temperature for hardening of membrane were studied. The sensing membrane, consisting of 27.5 wt.% of o-nitrophenyl octyl ether. 27.5 wt.% of tri-n-octylmethylammonium nitrate and 45 wt.% of raw Urushi latex, was coated with a thickness of 0.5 mm on a silver disc which was plated with Ag/AgCl, then plated with copper and hardened in the oven at 80 degrees C for 50 h. A semi-logarithmic calibration curve of potential versus nitrate ion concentration was obtained over the range 6-60 000 mg l(-1) NO(3)(poststaggered-). The slope of the linear part of the curve was -56 mV per decade change in NO(3)(poststaggered-) concentration. Compared with a PVC matrix nitrate ion-selective electrode, the Urushi matrix nitrate ion-selective electrode was superior in terms of hardness and mechanical strength of the membrane, short response time and long life. The combination of an Urushi matrix nitrate ion-selective electrode with a porous PTFE junction reference electrode, air-tight structured KCl solution chamber and a temperature sensor was applied to field measurements of nitrate-nitrogen concentrations in upland soils. The values obtained for upland soils containing 30-50% of water were good agreement with those for soil solution.

Entities:  

Year:  1996        PMID: 18966676     DOI: 10.1016/0039-9140(96)01963-7

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Solid-Contact Potentiometric Anion Sensing Based on Classic Silver/Silver Insoluble Salts Electrodes without Ion-Selective Membrane.

Authors:  Chunxian Liao; Lijie Zhong; Yitian Tang; Zhonghui Sun; Kanglong Lin; Longbin Xu; Yan Lyu; Dequan He; Ying He; Yingming Ma; Yu Bao; Shiyu Gan; Li Niu
Journal:  Membranes (Basel)       Date:  2021-12-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.