Literature DB >> 18966446

Correction of iron interface in the spectrophotometric flow injection catalytic determination of molybdenum in plants.

E N Carrilho1, F J Krug, E A Zagatto.   

Abstract

Iron interference in the spectrophotometric catalytic determination of molybdenum based on the iodide-hydrogen peroxide reaction can be corrected by using sulphosalicylic acid as masking and color-forming reagent. The catalytic influence of iron ions is circumvented to the extent of about 90% and correction of any remaining iron ions is possible by monitoring the colored iron(III)-salicylate complex at 490 nm. In this way, iron is also determined. With the proposed system, molybdenum can be determined in plant and food digests within the 0-100 mug Mo 1(-1) range in the presence of up to 25 mg Fe 1(-1), at a sampling rate of about 50 determinations h(-1). The relative standard deviation of 10 consecutive measurements was estimated as < 2%. Results for samples were comparable with those obtained by graphite furnace atomic absorption spectrometry. In addition, recoveries within the range 94-100% were calculated.

Entities:  

Year:  1995        PMID: 18966446     DOI: 10.1016/0039-9140(95)01688-0

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  A Highly Sensitive Multicommuted Flow Analysis Procedure for Photometric Determination of Molybdenum in Plant Materials without a Solvent Extraction Step.

Authors:  Felisberto G Santos; Boaventura F Reis
Journal:  J Anal Methods Chem       Date:  2017-03-05       Impact factor: 2.193

  1 in total