| Literature DB >> 18965991 |
O Fatibello-Filho1, J A Nóbrega, A J Moraes Guaritá-Santos.
Abstract
A flow injection potentiometric procedure for saccharin determination in dietary products is proposed. Saccharin is precipitated as mercurous saccharinate and the excess of the mercurous cation is potentiometrically measured using a silver wire coated with a mercury film as the working electrode. A filter unit is used to avoid contact between the precipitate and the electrode surface. With relocation in the flow manifold, the accumulated precipitate is removed on-line. Sucrose, glucose, aspartame, sodium cyclamate and sodium benzoate do not interfere when present in amounts similar to those observed in commercial products. Results are comparable with those obtained by UV-spectrophotometry and the correlation coefficient between methods is equal to 0.9930. A linear relationship between DeltaE (mV) and the logarithm of saccharin concentration was obtained in the saccharin concentration range 2 x 10(-3) - 1 x 10(-2)M. The sampling frequency is 60/hour and only 0.76 mg of Hg(2+)(2) is consumed in each determination.Entities:
Year: 1994 PMID: 18965991 DOI: 10.1016/0039-9140(93)e0018-9