Literature DB >> 18965597

Flow injection determination of chromium(III) by pyrogallol chemiluminescence.

S Nakano1, M Fukuda, S Kageyama, H Itabashi, T Kawashima.   

Abstract

A flow injection method is proposed for the determination of nanogram amounts of chromium(III) using a pyrogallol chemiluminescence system. It is based on its catalytic effect on the oxidation of pyrogallol with periodate at a neutral medium. The addition of 3-(N-morpholino)propanesulphonic acid to the reaction system increased the chemiluminescence signal for chromium(III). The present method allows the determination of 5-100ng/ml of chromium(III). The relative standard deviation of 2.2% (n = 10) was obtained at 20 ng/ml of chromium(III) and the detection limit (signal-to-noise ratio = 2) was 1 ng/ml with the sampling frequency of 25/hr.

Entities:  

Year:  1993        PMID: 18965597     DOI: 10.1016/0039-9140(93)80143-f

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057