| Literature DB >> 18965185 |
O Fatibello-Filho1, A A Suleiman, G G Guilbault.
Abstract
Formaldehyde in air was detected and assayed with a piezoelectric quartz crystal coated with a 7,10-dioxa-3,4-diaza-1,5,12,16-hexadecatetrol/chromotropic acid solution. Water vapor and several gaseous interferents were removed by passing the sampling stream through a column of anhydrous magnesium perchlorate. The response curves were linear in the concentration ranges 0.4-4.5 and 0.4-3.6 ppm v/v CH(2)O with and without the scrubber column, respectively. A single coating was used for 12 days (500 assays) without significant loss in sensitivity. With a single-point daily recalibration, the useful lifetime of the coating is about 2 months.Entities:
Year: 1991 PMID: 18965185 DOI: 10.1016/0039-9140(91)80178-3
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057