| Literature DB >> 18965161 |
Abstract
A precise and accurate method in which ammonia and ammonium ions are coulometrically titrated with electrogenerated hypobromite was studied. Through the measurement of the current efficiency for generating hypobromite and the dependencies of titration results on electrolysis current and delay time between sampling and starting electrolysis, precision and accuracy of 0.01 and 0.02%, respectively, were obtained. The accuracy of the titration was compared with that of acidimetric coulometric titration of ammonia.Entities:
Year: 1991 PMID: 18965161 DOI: 10.1016/0039-9140(91)80077-d
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057