Revision of failed knee replacements using fixed-axis hinges.

A E Inglis1, P S Walker.   

Abstract

Forty failed hinged arthroplasties of the knee were revised by the insertion of another hinged implant. In 14 cases the prosthesis used at the revision operation was similar to the primary implant; in 26, a hinge with an elongated femoral stem was used, usually replacing part of the femoral shaft. In seven of these knees an elongated tibial stem was also required, though the tibial shaft was replaced in only two of them. There were many complications. Fracture of the femur at the tip of the femoral stem was the most frequent. Sixteen first revisions failed and were revised a second time; 12 required replacement of the distal femoral shaft and three required replacement of the proximal tibia. The incidence of complications in knees requiring a second revision was even higher. Four required a third revision after an average interval of three years. Failure of a hinged prosthesis results in bone loss mainly in the femur. Revision of a failed hinged prosthesis with another of the same design is unlikely to be successful and may cause fracture of the femur.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1894661

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  14 in total

Review 1.  How to read a postoperative knee replacement radiograph.

Authors:  Nawfal Al-Hadithy; Madhavan C Papanna; Sana Farooq; Yegappan Kalairajah
Journal:  Skeletal Radiol       Date:  2011-10-16       Impact factor: 2.199

2.  Revision of broken knee megaprostheses: new solution to old problems.

Authors:  Manish Agarwal; Ashish Gulia; B Ravi; Rupesh Ghyar; Ajay Puri
Journal:  Clin Orthop Relat Res       Date:  2010-11       Impact factor: 4.176

Review 3.  Megaprostheses in the treatment of periprosthetic fractures of the knee joint: indication, technique, results and review of literature.

Authors:  Reinhard Windhager; Markus Schreiner; Kevin Staats; Sebastian Apprich
Journal:  Int Orthop       Date:  2015-09-25       Impact factor: 3.075

4.  Periprosthetic fractures around the knee: current concepts and advances in management.

Authors:  Michael R Whitehouse; Sanchit Mehendale
Journal:  Curr Rev Musculoskelet Med       Date:  2014-06

5.  Outcome of osteosynthesis for periprosthetic fractures after total knee arthroplasty: a retrospective study.

Authors:  Hasmukh Nagwadia; Prateek Joshi
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-01-03

6.  Primary total knee arthroplasty using rotating-hinge prosthesis in severely affected knees.

Authors:  Jae-Hyuk Yang; Jung-Ro Yoon; Chi-Hun Oh; Taik-Sun Kim
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-07-20       Impact factor: 4.342

7.  Compress periprosthetic fractures: interface stability and ease of revision.

Authors:  Wakenda K Tyler; John H Healey; Carol D Morris; Patrick J Boland; Richard J O'Donnell
Journal:  Clin Orthop Relat Res       Date:  2009-06-30       Impact factor: 4.176

8.  Results of a Second-generation Constrained Condylar Prosthesis in Complex Primary and Revision Total Knee Arthroplasty: A Mean 5.5-Year Follow-up.

Authors:  Chen-Yi Ye; De-Ting Xue; Shuai Jiang; Rong-Xin He
Journal:  Chin Med J (Engl)       Date:  2016-06-05       Impact factor: 2.628

9.  Mid-term results after implantation of rotating-hinge knee prostheses: primary versus revision.

Authors:  Turgay Efe; Philip P Roessler; Thomas J Heyse; Carsten Hauk; Caroline Pahrmann; Alan Getgood; Jan Schmitt
Journal:  Orthop Rev (Pavia)       Date:  2012-12-11

10.  Periprosthetic fractures around the knee-the best way of treatment.

Authors:  Steffen Ruchholtz; Jordi Tomás; Florian Gebhard; Morten Schultz Larsen
Journal:  Eur Orthop Traumatol       Date:  2012-08-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.