Twenty years of single center experience in ESWL 1987-2007: an evaluation of 3079 patients.

Geert G Tailly1, Joost A Baert, K Robert Hente, Thomas O Tailly.   

Abstract

INTRODUCTION: Since 1987 we have consecutively operated five different Dornier lithotriptors in the management of lithiasis at all levels of the urinary tract. At the time of usage of each machine we periodically evaluated the treatment results in order to eventually adapt treatment strategies to improve results. None of these evaluations comprised the totality of patients treated on a certain machine during the entire period it was in use. This overview retrospectively compares the results over the entire period in order to gauge the impact of the different machines and SW-sources on clinical results.
MATERIAL AND METHODS: A total of 3079 patients, treated from september 1987 till september 2006 with two electrohydraulic machines, HM4 (1987-1988) and MPL9000 (1988-1994), and three electromagnetic machines, Compact (1991-1999), DoLi U/50 (1996-1999) and DoLi S (1999-2006) were thus evaluated. Parameters reviewed were stone location and size, retreatment rate, auxiliary procedures, stone free rate and Effectiveness Quotient (EQ).
RESULTS: There is no significant difference in stone size or stone free rate (p=0.4715) for the different groups. There is a significant difference for auxiliary procedure rate and retreatment rate for the different machines. The DoLiS with EMSE 220F-XXP scores significantly better than any of the other systems for auxiliary procedure rate pre-ESWL (p<0.05) and total auxiliary procedure rate (p<0.05), and retreatment rate (p=0.0024). There is a gradual increase in EQ with the consecutive machines.
CONCLUSION: Although stone free rates (85%-88.8%) remained constant, Effectiveness Quotients improved. This is largely due to reduced auxiliary procedure rates and reduced retreatment rates. The first effect is a consequence of improvement in treatment strategies and experience, the latter no doubt also of improvement in SW-sources.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18937585     DOI: 10.1089/end.2008.9706

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  4 in total

1.  Strategic lithotripsy using the Doli S EMSE 220 F-XP for the management of staghorn renal calculi.

Authors:  Ioannis Heretis; Charalampos Mamoulakis; Vaios Papadimitriou; Frank Sofras
Journal:  Int Urol Nephrol       Date:  2010-06-08       Impact factor: 2.370

Review 2.  Arguments for choosing extracorporeal shockwave lithotripsy for removal of urinary tract stones.

Authors:  Hans-Göran Tiselius; Christian G Chaussy
Journal:  Urolithiasis       Date:  2015-08-28       Impact factor: 3.436

3.  How effective is extracorporeal shock wave lithotripsy of ureteral stones with Dornier Lithotripter S EMSE 220F-XXP? A prospective and preliminary assessment.

Authors:  Maria Chiara Sighinolfi; Salvatore Micali; Stefano De Stefani; Giovanni Alberto Pini; Massimo Rivalta; Filippo Cianci; Giampaolo Bianchi
Journal:  Surg Endosc       Date:  2010-09-16       Impact factor: 4.584

4.  Focused ultrasound to expel calculi from the kidney: safety and efficacy of a clinical prototype device.

Authors:  Jonathan D Harper; Mathew D Sorensen; Bryan W Cunitz; Yak-Nam Wang; Julianna C Simon; Frank Starr; Marla Paun; Barbrina Dunmire; H Denny Liggitt; Andrew P Evan; James A McAteer; Ryan S Hsi; Michael R Bailey
Journal:  J Urol       Date:  2013-04-09       Impact factor: 7.450

  4 in total