Elimination of matrix and spectral interferences in the measurement of lead and cadmium in urine and blood by electrothermal atomic absorption spectrometry with deuterium background correction.

P C D'Haese1, L V Lamberts, L Liang, F L Van de Vyver, M E De Broe.   

Abstract

Direct measurement of lead and cadmium in blood and urine by electrothermal atomic absorption spectrometry with deuterium background correction (D2-AAS) is prone to severe matrix and spectral interferences. We overcame these effects by coating the L'vov platform with ammonium molybdate, reducing the atomization time, introducing a post-atomization cooling step, carefully selecting ashing and atomization temperatures, and using an appropriate procedure for matrix modification. To determine Pb and Cd in blood and urine, we used matrix-matched calibration curves. With the proposed procedure for sample preparation, both Pb and Cd in whole blood can be determined in the same diluted sample. Results obtained by D2-AAS correlate closely with those by Zeeman-corrected AAS. Detection limits (mean blank + 3 SDblank) for Pb in urine and blood were 4 micrograms/L. For cadmium, the detection limits were 0.4 and 0.1 micrograms/L for urine and blood analysis, respectively. Between-run CVs were less than 5.0%.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1893594

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  Determination of cadmium by electrothermal atomic absorption spectrometry after microwave-assisted digestion of animal tissues and sewage sludges.

Authors:  R Chakraborty; A K Das; M L Cervera; M De La Guardia
Journal:  Anal Bioanal Chem       Date:  1996-04       Impact factor: 4.142

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.