Cementless Gustilo-Kyle and BIAS total hip arthroplasty: 2- to 5-year results.

H Kienapfel1, J Martell, A Rosenberg, J Galante.   

Abstract

We conducted extensive radiographic and clinical examination of 58 cementless total hip arthroplasties. Forty BIAS femoral stems had heads made of Co-base F 75 alloy and 18 Gustilo-Kyle femoral stems had Ti6A14V alloy heads. The cross-sectional geometry and location of the porous coating pads of the two stem types was identical, the BIAS component being slightly shorter. At the last follow-up, 27.8% of the Gustilo-Kyle femoral components and 2.5% of the BIAS femoral components were unstable. None of the acetabular components in the BIAS group as against 38.9% of the acetabular components in the Gustilo-Kyle group demonstrated wear of more than 1 mm. All unstable components which did not suffer intraoperative fracture had also acetabular wear of more than 1 mm. The only case with endosteal erosion was seen in a patient with an unstable implant displaying the most extensive wear. Nonparametric statistical analysis showed that the patients with Ti6A14V/polyethylene bearing surfaces had a statistically significantly higher extent of wear and a statistically significant higher incidence of femoral component loosening than patients with CrCoMo/polyethylene bearing surfaces.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1892712     DOI: 10.1007/bf00571055

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  30 in total

1.  Can wear in total hip arthroplasties be assessed from radiographs?

Authors:  J C Clarke; K Black; C Rennie; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1976 Nov-Dec       Impact factor: 4.176

2.  Biomechanical investigations of bone-porous carbon and porous metal interfaces.

Authors:  J L Nilles; M Lapitsky
Journal:  J Biomed Mater Res       Date:  1973

3.  The long-term results of low-friction arthroplasty of the hip performed as a primary intervention.

Authors:  J Charnley
Journal:  J Bone Joint Surg Br       Date:  1972-02

4.  Sintered fiber metal composites as a basis for attachment of implants to bone.

Authors:  J Galante; W Rostoker; R Lueck; R D Ray
Journal:  J Bone Joint Surg Am       Date:  1971-01       Impact factor: 5.284

5.  Cemented total hip replacement in patients who are less than fifty years old.

Authors:  D K Collis
Journal:  J Bone Joint Surg Am       Date:  1984-03       Impact factor: 5.284

6.  Ectopic ossification following total hip replacement. Incidence and a method of classification.

Authors:  A F Brooker; J W Bowerman; R A Robinson; L H Riley
Journal:  J Bone Joint Surg Am       Date:  1973-12       Impact factor: 5.284

7.  "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

Authors:  T A Gruen; G M McNeice; H C Amstutz
Journal:  Clin Orthop Relat Res       Date:  1979-06       Impact factor: 4.176

8.  15-21-year results of the Charnley low-friction arthroplasty.

Authors:  B M Wroblewski
Journal:  Clin Orthop Relat Res       Date:  1986-10       Impact factor: 4.176

9.  Socket wear in Charnley low friction arthroplasty of the hip.

Authors:  M J Griffith; M K Seidenstein; D Williams; J Charnley
Journal:  Clin Orthop Relat Res       Date:  1978 Nov-Dec       Impact factor: 4.176

10.  Total hip arthroplasties in patients less than forty-five years old.

Authors:  L D Dorr; G K Takei; J P Conaty
Journal:  J Bone Joint Surg Am       Date:  1983-04       Impact factor: 5.284

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.