Computer-assisted distraction osteogenesis by Ilizarov's method.

A L Simpson1, B Ma, B Slagel, D P Borschneck, R E Ellis.   

Abstract

BACKGROUND: The Taylor spatial frame is a fixation device used to implement the Ilizarov method of bone deformity correction to gradually distract an osteotomized bone at regular intervals, according to a prescribed schedule.
METHODS: We modify conventional technique by: (a) preoperatively planning a virtual three-dimensional (3D) correction; (b) basing the correction on the actual location of the frame with respect to the anatomy, immediately compensating for frame mounting errors; and (c) calculating the correction based on 3D CT data rather than measurements from radiographs. We have performed a laboratory study using plastic phantoms, and a pilot clinical study involving five patients.
RESULTS: In 20 tibial phantom experiments, we achieved average correction errors of < 2 degrees total rotation and < 0.5 mm total lengthening. We observed clinically acceptable corrections with no complications in our pilot clinical study.
CONCLUSIONS: Our method achieved high accuracy and precision in a laboratory setting, and produced acceptable outcomes in a pilot clinical study.

Entities:  

Mesh:

Year:  2008        PMID: 18924116     DOI: 10.1002/rcs.211

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  5 in total

1.  Patient-tailored plate for bone fixation and accurate 3D positioning in corrective osteotomy.

Authors:  J G G Dobbe; J C Vroemen; S D Strackee; G J Streekstra
Journal:  Med Biol Eng Comput       Date:  2012-10-10       Impact factor: 2.602

2.  Configuration design and correction ability evaluation of a novel external fixator for foot and ankle deformity treated by U osteotomy.

Authors:  Shiping Zuo; Mingjie Dong; Jianfeng Li; Chunjing Tao; Run Ji
Journal:  Med Biol Eng Comput       Date:  2020-01-08       Impact factor: 2.602

3.  Computer-assisted oblique single-cut rotation osteotomy to reduce a multidirectional tibia deformity: case report.

Authors:  J G G Dobbe; K J du Pré; L Blankevoort; G J Streekstra; P Kloen
Journal:  Strategies Trauma Limb Reconstr       Date:  2017-03-01

4.  Marker- three dimensional measurement versus traditional radiographic measurement in the treatment of tibial fracture using Taylor spatial frame.

Authors:  Qixin Liu; Yanshi Liu; Hong Li; Xuefei Fu; Xingpeng Zhang; Sida Liu; Jinli Zhang; Tao Zhang
Journal:  BMC Musculoskelet Disord       Date:  2022-02-16       Impact factor: 2.362

5.  Computer-assisted preoperative simulation for positioning and fixation of plate in 2-stage procedure combining maxillary advancement by distraction technique and mandibular setback surgery.

Authors:  Hideyuki Suenaga; Asako Taniguchi; Kazumichi Yonenaga; Kazuto Hoshi; Tsuyoshi Takato
Journal:  Int J Surg Case Rep       Date:  2016-10-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.