Literature DB >> 1888037

Detection apparatus for multiple heterogeneous chemiluminescence immunoassay configurations.

O S Khalil1, T F Zurek, C Pepe, K Genger, D G Huff, C Coleman, C Hanna, R Hu, J Mackowiack, L Bennett.   

Abstract

We describe an apparatus for measuring signals emanated from two heterogeneous chemiluminescence immunoassay (CLIA) configurations: antibody-coated polystyrene beads, in reaction tray wells, and microparticles captured by a porous matrix. An optics and fluidics design which allows the use of a common detection head for these two different assay configurations is described. The detection head moves along three Cartesian coordinates to create a localized light-tight compartment around each individual disposable reaction vessel. Reproducibility of the light seal, trigger solution delivery, and mixing is achieved for acridinium-labeled CLIA. The coated polystyrene beads configuration is tested using beta HCG, CEA, and TSH assays. The microparticle-capture configuration is tested using beta HCG and HBsAg assays. The microparticle capture CLIA has shorter incubation times and the potential for ease of automation.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1888037     DOI: 10.1016/0003-2697(91)90117-c

Source DB:  PubMed          Journal:  Anal Biochem        ISSN: 0003-2697            Impact factor:   3.365


  1 in total

1.  Annual periodic increases in serum carcinoembryonic antigen concurrent with ground-glass opacity in the lung: report of a case.

Authors:  Tsutomu Sakuma; Yumiko Iwata; Yoshimichi Ueda; Xiu Gu; Makoto Sugita; Motoyasu Sagawa
Journal:  Surg Today       Date:  2005       Impact factor: 2.549