| Literature DB >> 1885682 |
Abstract
A commercially available system is described for the fully automated clean-up and high-performance liquid chromatographic (HPLC) analysis of aflatoxins in foods and animal feeds. The system marketed primarily for handling solid-phase extraction columns has modified software to facilitate use with immunoaffinity columns. Sample extract clean-up followed by injection onto an HPLC column with post-column iodination and fluorescence detection is carried out completely unattended. A coefficient of variation of 5.1% for aflatoxin B1 analysis was obtained, and the accuracy of the system was demonstrated by the analysis of peanut butter certified reference material.Entities:
Mesh:
Substances:
Year: 1991 PMID: 1885682 DOI: 10.1016/s0021-9673(01)95771-1