Literature DB >> 18853285

Safety and feasibility of percutaneous closure of patent foramen ovale without intra-procedural echocardiography in 825 patients.

Andreas Wahl1, Fabien Praz, Jessica Stinimann, Stephan Windecker, Christian Seiler, Krassen Nedeltchev, Heinrich P Mattle, Bernhard Meier.   

Abstract

BACKGROUND: Percutaneous closure of patent foramen ovale (PFO) is generally performed using intra-procedural guidance by transoesophageal (TEE) or intracardiac (ICE) echocardiography. While TEE requires sedation or general anaesthesia, ICE is costly and adds incremental risk, and both imaging modalities lengthen the procedure.
METHODS: A total of 825 consecutive patients (age 51 +/- 13 years; 58% male) underwent percutaneous PFO closure solely under fluoroscopic guidance, without intra-procedural echocardiography. The indications for PFO closure were presumed paradoxical embolism in 698 patients (95% cerebral, 5% other locations), an embolic event with concurrent aetiologies in 47, diving in 51, migraine headaches in 13, and other reasons in 16. An atrial septal aneurysm was associated with the PFO in 242 patients (29%).
RESULTS: Permanent device implantation failed in two patients (0.2%). There were 18 procedural complications (2.2%), including embolization of the device or parts of it in five patients with successful percutaneous removal in all cases, air embolism with transient symptoms in four patients, pericardial tamponade requiring pericardiocentesis in one patient, a transient ischaemic attack with visual symptoms in one patient, and vascular access site problems in seven patients. There were no long-term sequelae. Contrast TEE at six months showed complete abolition of right-to-left shunt via PFO in 88% of patients, whereas a minimal, moderate or large residual shunt persisted in 7%, 3%, and 2%, respectively.
CONCLUSIONS: This study confirms the safety and feasibility of percutaneous PFO closure without intra-procedural echocardiographic guidance in a large cohort of consecutive patients.

Entities:  

Mesh:

Year:  2008        PMID: 18853285     DOI: 2008/39/smw-12255

Source DB:  PubMed          Journal:  Swiss Med Wkly        ISSN: 0036-7672            Impact factor:   2.193


  3 in total

1.  Does persistent (patent) foramen ovale closure reduce the risk of recurrent decompression sickness in scuba divers?

Authors:  Björn Edvinsson; Ulf Thilén; Niels Erik Nielsen; Christina Christersson; Mikael Dellborg; Peter Eriksson; Joanna Hlebowicz
Journal:  Diving Hyperb Med       Date:  2021-03-31       Impact factor: 0.887

2.  Feasibility and outcomes of combined transcatheter aortic valve replacement with other structural heart interventions in a single session: a matched cohort study.

Authors:  Ahmed A Khattab; Steffen Gloekler; Beate Sprecher; Samera Shakir; Enio Guerios; Stefan Stortecky; Crochan J O'Sullivan; Fabian Nietlispach; Aris Moschovitis; Thomas Pilgrim; Lutz Buellesfeld; Peter Wenaweser; Stephan Windecker; Bernhard Meier
Journal:  Open Heart       Date:  2014-06-27

3.  Multidisciplinary assessment of PFO with substantial right-to-left shunting and medium-term follow-up after PFO device closure: A single-center experience.

Authors:  Lu He; Ge-Sheng Cheng; Ya-Juan Du; Yu-Shun Zhang
Journal:  J Interv Cardiol       Date:  2017-06-01       Impact factor: 2.279