Literature DB >> 18850547

Topography-guided treatment of decentered laser ablation using LaserSight's excimer laser.

L Wu1, X Zhou, Z Ouyang, C Weng, R Chu.   

Abstract

PURPOSE: To assess the efficacy of topography-guided laser ablation for correction of previously decentered laser ablation using LaserSight's excimer laser.
METHODS: Re-treatment was performed to correct decentered ablation using LaserSight's excimer laser for 18 patients who previously underwent LASIK surgery for myopia correction in both eyes. For each patient, only the decentered eye was re-treated while the other asymptomatic eye forms a control group for this study. Measurements were conducted on ablation center, best spectacle-corrected visual acuity (BSCVA), contrast sensitivity and corneal aberrations pre- and post-operatively.
RESULTS: For the retreated 18 eyes, the mean decentration was significantly reduced from 1.32+/-0.28mm to 0.61+/-0.23mm post-operatively (t=16.24, p<0.001), and with a significant improvement in mean BSCVA from 0.08+/-0.09 logMAR to 0.01+/-0.11 logMAR (t=4.58, p<0.001). The post-operative contrast sensitivity at the spatial frequencies (SF) of 1.00 and 0.70 was significantly improved (p<0.05 for both SFs). Corneal higher-order aberrations (HOAs), including the coma-like aberrations and spherical aberration, were decreased. In comparing the measurements for the retreated group to those for the control group, no significant differencewas found either in decentration or in BSCVA, but the contrast sensitivity at 0.70 was lower and the level of corneal aberrations was higher.
CONCLUSIONS: Topography-guided ablation with LaserSight excimer laser is effective to correct decentered ablation. However, the re-treated eye is still inferior to the eye with originally centered ablation in corneal optical quality or visual performance.

Entities:  

Mesh:

Year:  2008        PMID: 18850547     DOI: 10.1177/112067210801800508

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  2 in total

1.  Using LaserSight Astrapro Planner 2.2 Z software in corneal topography-guided laser in situ keratomileusis for myopia with asymmetric corneal shape.

Authors:  Bing Liu; Wei Chen; De-Wang Shao; Hua Wang; Hai-Xia Ru; Min Zhang; Ying Wang; Chun-Yan Yang
Journal:  Int J Ophthalmol       Date:  2014-06-18       Impact factor: 1.779

Review 2.  Bilateral symmetry in vision and influence of ocular surgical procedures on binocular vision: A topical review.

Authors:  Samuel Arba Mosquera; Shwetabh Verma
Journal:  J Optom       Date:  2016-03-16
  2 in total