Literature DB >> 18843057

Can you deliver accurate tidal volume by manual resuscitator?

H M Lee1, K H Cho, Y H Choi, S Y Yoon, Y H Choi.   

Abstract

OBJECTIVES: One of the problems with manual resuscitators is the difficulty in achieving accurate volume delivery. The volume delivered to the patient varies by the physical characteristics of the person and method. This study was designed to compare tidal volumes delivered by the squeezing method, physical characteristics and education and practice levels.
METHODS: 114 individuals trained in basic life support and bag-valve-mask ventilation participated in this study. Individual characteristics were obtained by the observer and the education and practice level were described by the subjects. Ventilation was delivered with a manual resuscitator connected to a microspirometer and volumes were measured. Subjects completed three procedures: one-handed, two-handed and two-handed half-compression.
RESULTS: The mean (standard deviation) volumes for the one-handed method were 592.84 ml (SD 117.39), two-handed 644.24 ml (SD 144.7) and two-handed half-compression 458.31 ml (SD 120.91) (p<0.01). Tidal volume delivered by two hands was significantly greater than that delivered by one hand (r = 0.398, p<0.01). The physical aspects including hand size, volume and grip power had no correlation with the volume delivered. There were slight increases in tidal volume with education and practice, but correlation was weak (r = 0.213, r = 0.281, r = 0.131, p<0.01).
CONCLUSIONS: The tidal volume delivered by a manual resuscitator shows large variations. There were significant differences in the volume delivered by compression methods, but physical characteristics are not a predictor of tidal volume delivery. The manual resuscitator is not a suitable device for accurate ventilation.

Entities:  

Mesh:

Year:  2008        PMID: 18843057     DOI: 10.1136/emj.2007.053678

Source DB:  PubMed          Journal:  Emerg Med J        ISSN: 1472-0205            Impact factor:   2.740


  9 in total

1.  Effect of the laryngeal tube on the no-flow-time in a simulated two rescuer basic life support setting with inexperienced users.

Authors:  J Schröder; M Bucher; O Meyer
Journal:  Med Klin Intensivmed Notfmed       Date:  2015-09-15       Impact factor: 0.840

2.  Effect of metronome rates on the quality of bag-mask ventilation during metronome-guided 30:2 cardiopulmonary resuscitation: A randomized simulation study.

Authors:  Ji Ung Na; Sang Kuk Han; Pil Cho Choi; Dong Hyuk Shin
Journal:  World J Emerg Med       Date:  2017

3.  Effect of using a laryngeal tube on the no-flow time in a simulated, single-rescuer, basic life support setting with inexperienced users.

Authors:  O Meyer; M Bucher; J Schröder
Journal:  Anaesthesist       Date:  2016-02-17       Impact factor: 1.041

4.  Advantages of the New Tao Mask for Bag Mask Ventilation: a randomized crossover trial.

Authors:  Catherine D Tobin; Tamas A Szabo; Bethany J Wolf; Kathryn H Bridges; Tod A Brown; Erick M Woltz; Robert D Warters
Journal:  Rom J Anaesth Intensive Care       Date:  2018-10

5.  Standard operating procedure changed pre-hospital critical care anaesthesiologists' behaviour: a quality control study.

Authors:  Leif Rognås; Troels Martin Hansen; Hans Kirkegaard; Else Tønnesen
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2013-12-05       Impact factor: 2.953

6.  Evaluation of Bag-Valve-Mask Ventilation in Manikin Studies: What Are the Current Limitations?

Authors:  A Khoury; F S Sall; A De Luca; A Pugin; S Pili-Floury; L Pazart; G Capellier
Journal:  Biomed Res Int       Date:  2016-05-16       Impact factor: 3.411

7.  Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill Patients: A Randomised, Controlled, Cross-Over Simulation Study.

Authors:  Jong Won Kim; Sang O Park; Kyeong Ryong Lee; Dae Young Hong; Kwang Je Baek
Journal:  Emerg Med Int       Date:  2020-04-24       Impact factor: 1.112

8.  An Emergency Amendment to the National Scope of Practice for Paramedics in the Setting of a Global Pandemic.

Authors:  Brad Boehringer; Peter O'Meara; Gary Wingrove; Nikiah G Nudell
Journal:  J Rural Health       Date:  2020-06-25       Impact factor: 5.667

9.  Tidal volume according to the 4-point sealing forces of a bag-valve-mask: an adult respiratory arrest simulator-based prospective, descriptive study.

Authors:  Dongchoon Uhm; Ajung Kim
Journal:  BMC Emerg Med       Date:  2021-05-01
  9 in total